Notifications
You must be signed in to change notification settings - Fork 32
Open
Labels
bugSomething isn't workingSomething isn't working
Description
Hi!
Im using your code and the problem i have is that my results in IR and RED data are somehow swapped i guess? I mean, in logs you've uploaded IR result is around 20k higher than RED one, but when i take a measure I get higher values in RED data. I observed it when i tried to get spo value manually by dividing ac and dc currents. When i inverted R factor i got a proper value almost everytime. I also check the MAX30102 modes. In hr mode only red light should be turned on, in spo mode both should be turned on. But when i turn the hr mode i see no light. The light apperas only in spo mode. This fact would confirm that signals are inverted or something i guess? Because red light is visible and ir is not for us. What do you think?
Metadata
Metadata
Assignees
Labels
bugSomething isn't workingSomething isn't working